Kelvin probe as an effective tool to develop sensitive p-type CuO gas sensors
Year: 2016
Authors: Mazhar M.E., Faglia G., Comini E., Zappa D., Baratto C., Sberveglieri G.
Autors Affiliation: Univ Brescia, SENSOR, Dipartimento Ingn Informaz, I-25123 Brescia, Italy; CNR, INO, UOS Brescia, I-25123 Brescia, Italy.
Abstract: We present a comparative study of gas sensing properties of p-type CuO nanowires, characterized by Kelvin Probe (KP) surface workfunction gas sensing and formal conductometric (chemoresistive) gas sensing schemes, with main emphasis on the detection performance of low concentrations of butanol. P-type CuO nanowires were prepared by thermal oxidation of Cu thin film in a controlled environment at oxidation temperatures of 300 degrees C and 400 degrees C. Structural characterizations of grown nanomaterials were carried out by X-ray diffraction, micro-Raman spectrophotometry and scanning electron microscopy (SEM). Butanol sensing responses recorded from both methods were compared and further analyzed to calculate approximate values of electron affinity variations.
Journal/Review: SENSORS AND ACTUATORS B-CHEMICAL
Volume: 222 Pages from: 1257 to: 1263
More Information: This work is partly supported by European project “ORAMA “Oxide Materials Towards a Mature Post-silicon Electronics Era”, NMP-2009-2.2.1, “WIROX” Oxide Nanostructures for Wireless Chemical Sensing, PEOPLE MARIE CURIE ACTIONS, International Research Staff Exchange Scheme, Call: FP7-PEOPLE-2011-IRSES, 2012-2015, as well as by European CommunityKeyWords: Butanol sensing; Conductometric sensing; CuO nanowires; Electron affinity; Kelvin probe; P-type metal oxides; Surface workfunctionDOI: 10.1016/j.snb.2015.05.050ImpactFactor: 5.401Citations: 36data from “WEB OF SCIENCE” (of Thomson Reuters) are update at: 2024-10-27References taken from IsiWeb of Knowledge: (subscribers only)Connecting to view paper tab on IsiWeb: Click hereConnecting to view citations from IsiWeb: Click here