Differential microcalorimeter for liquid samples
Anno: 1989
Autori: Barbini A., Bertolini D., Cassettari M., Papucci F., Salvetti A., Salvetti G., Veronesi S.
Affiliazione autori: Istituto di Fisica Atomica e Molecolare del C.N.R., via del Giardino 7, Pisa, Italy
Abstract: A microcalorimeter for liquid samples with high performance and low cost is described. It is essentially a differential apparatus in a \”twin\” configuration with two cylindrical cells. Two resistive sensors, put in a Wheatstone bridge, are wound on the cells to measure the temperature difference between them. Sensitivity, response time, and calibration procedure are reported. As illustrative example, the specific heat of some liquids at two temperatures has been measured.
Giornale/Rivista: REVIEW OF SCIENTIFIC INSTRUMENTS
Volume: 60 (7) Da Pagina: 1308 A: 1311
Parole chiavi: Thermodynamic measurements and instrumentation; Thermodynamic properties; Wheatstone bridge circuit DOI: 10.1063/1.1140982Citazioni: 12dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2024-12-08Riferimenti tratti da Isi Web of Knowledge: (solo abbonati) Link per visualizzare la scheda su IsiWeb: Clicca quiLink per visualizzare la citazioni su IsiWeb: Clicca qui