Characterization and calibration of a variable-angle absolute reflectometer
Anno: 1990
Autori: Castellini C., Emiliani G., Masetti E., Poggi P., Polato P.P.
Affiliazione autori: Comitato Nazionale per la Ricerca e lo Sviluppo dell’Energia Nucleare e delle Energie Alternative, S.P. Anguillarese Km. 1300-Casaccia-Roma, Italy; Istituto Nazionale di Ottica, Largo E. Fermi 6, 50125 Arcetri-Firenze, Italy; Stazione sperimentale del Vetro, 10 Via Briati, Murano-Venezia, Italy
Abstract: The instrument described is a newly designed reflectometer. Measurements are made with a polarized light beam as a function of wavelengt hand angle of incidence. Calibration tests have outlined the possibility of using the instrument for reflectance measurements on mirrors with good reproducibility (0.2%) and accuracy (better than 1%). This has been obtained by the alignment system of the sample and the absolute method used for the determination of its reflectance .
Giornale/Rivista: APPLIED OPTICS
Volume: 29 (4) Da Pagina: 538 A: 543
Parole chiavi: reflectometer; calibration; DOI: 10.1364/AO.29.000538Citazioni: 11dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2025-03-16Riferimenti tratti da Isi Web of Knowledge: (solo abbonati) Link per visualizzare la scheda su IsiWeb: Clicca quiLink per visualizzare la citazioni su IsiWeb: Clicca qui