Ionizing Radiation Detectors Based on Ge-Doped Optical Fibers Inserted in Resonant Cavities

Anno: 2015

Autori: Avino S., D’Avino V., Giorgini A., Pacelli R., Liuzzi R., Cella L., De Natale P., Gagliardi G.

Affiliazione autori: Consiglio Nazionale delle Ricerche, Istituto Nazionale di Ottica (INO), via Campi Flegrei 34 — Comprensorio A. Olivetti, 80078 Pozzuoli (Na), Italy; Consiglio Nazionale delle Ricerche, Istituto di Biostrutture e Bioimmagini, via Tommaso De Amicis 95, 80131 Napoli, Italy; Università di Napoli Federico II, Dipartimento di Scienze Biomediche Avanzate, via Pansini 5, 80131 Napoli, Italy; Consiglio Nazionale delle Ricerche, Istituto Nazionale di Ottica (INO), Largo Enrico Fermi 6, 50125 Firenze, Italy

Abstract: The measurement of ionizing radiation (IR) is a crucial issue in different areas of interest, from environmental safety and industrial monitoring to aerospace and medicine. Optical fiber sensors have recently proven good candidates as radiation dosimeters. Here we investigate the effect of IR on germanosilicate optical fibers. A piece of Ge-doped fiber enclosed between two fiber Bragg gratings (FBGs) is irradiated with gamma radiation generated by a 6 MV medical linear accelerator. With respect to other FBG-based IR dosimeters, here the sensor is only the bare fiber without any special internal structure. A near infrared laser is frequency locked to the cavity modes for high resolution measurement of radiation induced effects on the fiber optical parameters. In particular, we observe a variation of the fiber thermo-optic response with the radiation dose delivered, as expected from the interaction with Ge defect centers, and demonstrate a detection limit of 360 mGy. This method can have an impact in those contexts where low radiation doses have to be measured both in small volumes or over large areas, such as radiation therapy and radiation protection, while bare optical fibers are cheap and disposable.

Giornale/Rivista: SENSORS

Volume: 15 (2)      Da Pagina: 4242  A: 4252

Maggiori informazioni: The authors acknowledge support by Italian Ministry for Education and Research (MIUR) with the project FIRB RBFR10Q0PT_001 \”DROPS\”. The authors acknowledge Gianluca Notariale and Andrea Finizio for technical support and Giuseppe Palma for helpful discussions.
Parole chiavi: Accident prevention; Dosimeters; Dosimetry; Fiber optic sensors; Fibers; Germanium; Infrared devices; Infrared lasers; Ionizing radiation; Optical fibers; Optical materials; Optical resonators; Radiation; Radiation detectors; Radiation effects; Radiation protection; Radiation shielding, Environmental safety; Fiber Bragg gratings (FBGs); High-resolution measurements; Industrial monitoring; Near-infrared lasers; Optical fiber sensor; Radiation dosimeters; Radiation-induced effects, Fiber Bragg gratings, glass fiber, fiber optics; ionizing radiation; procedures, Fiber Optic Technology; Optical Fibers; Radiation, Ionizing
DOI: 10.3390/s150204242

Citazioni: 6
dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2024-03-24
Riferimenti tratti da Isi Web of Knowledge: (solo abbonati)
Link per visualizzare la scheda su IsiWeb: Clicca qui
Link per visualizzare la citazioni su IsiWeb: Clicca qui