Characterization of instrumental line shape distortions due to path difference dependent phase errors in a Fourier transform spectrometer
Anno: 2000
Autori: Bianchini G., Raspollini P.
Affiliazione autori: Ist. Ric. sulle Onde E., Gruppo Stratosfera, Via P., Florence, Italy; Department of Physics, Univ. Studi Firenze, Largo E. F., Florence, Italy
Abstract: In Fourier transform spectroscopy, path difference dependent misalignments, such as those due to imperfect mechanical/movement of the scanning mirror, can be modelled as path difference dependent apodization and phase errors. The theoretical model that relates the mirror drive characteristics to the path difference dependent error is used to verify the consistency of the optical instrument line shape with the measurements of the mechanical drive quality. This procedure can be used as a diagnostic and qualification tool.
Giornale/Rivista: INFRARED PHYSICS & TECHNOLOGY
Volume: 41 (5) Da Pagina: 287 A: 292
Parole chiavi: Error analysis; Fourier transform infrared spectroscopy; Mathematical models, Instrumental line shape (ILS); Phase errors, Infrared spectrometersDOI: 10.1016/S1350-4495(00)00048-7Citazioni: 10dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2025-05-18Riferimenti tratti da Isi Web of Knowledge: (solo abbonati) Link per visualizzare la scheda su IsiWeb: Clicca quiLink per visualizzare la citazioni su IsiWeb: Clicca qui