A Molecule-Based Single-Photon Source Applied in Quantum Radiometry

Year: 2020

Authors: Lombardi P., Trapuzzano M., Colautti M., Margheri G., Degiovanni I.P., López M., Kück S., Toninelli C.

Autors Affiliation: Istituto Nazionale di Ottica (CNR-INO) via N. Carrara 1, Sesto Fiorentino 50019, Florence, Italy; Uversità degli Studi di Firenze via G. Sansone 1, Sesto Fiorentino 50019, Florence, Italy; LENS, Università degli Studi di Firenze
via N. Carrara 1, Sesto Fiorentino 50019, Florence, Italy; Istituto dei Sistemi Complessi (CNR-ISC) via Madonna del Piano 10, Sesto Fiorentino 50019, Florence, Italy; Physikalisch-Technische Bundesanstalt (PTB) Bundesallee 100, 38116 Braunschweig, Germany; Istituto Nazionale di Ricerca Metrologica (INRiM) Strada delle Caccie 91, 10135 Torino, Italy

Abstract: Single-photon sources (SPSs) based on quantum emitters hold promise in quantum radiometry as metrology standard for photon fluxes at the low light level. Ideally this requires control over the photon flux in a wide dynamic range, sub-Poissonian photon statistics, and narrow-band emission spectrum. In this work, a monochromatic SPS based on an organic dye molecule is presented, whose photon flux is traceably measured to be adjustable between 144 000 and 1320 000 photons per second at a wavelength of (785.6 ± 0.1) nm, corresponding to an optical radiant flux between 36.5 and 334 fW. The high purity of the single-photon stream is verified, with a second-order autocorrelation function at zero time delay below 0.1 throughout the whole range. Such molecule-based SPS is hence used for the calibration of a single-photon avalanche detector against a low-noise analog photodiode traceable to the primary standard for optical radiant flux (i.e., the cryogenic radiometer). Due to the narrow bandwidth of the source, corrections to the detector efficiency arising from the spectral power distribution are negligible. With this major advantage, the developed device may finally realize a low-photon-flux standard source for quantum radiometry.

Journal/Review: JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS

Volume: 3 (2)      Pages from: 1900083-1  to: 1900083-8

KeyWords: quantum radiometry, single molecules, single-photon detectors, single- photon sources
DOI: 10.1002/qute.201900083