A wearable sweat rate sensor to monitor the athletes’ performance during training

Anno: 2018

Autori: Salvo P., Pingitore A., Barbini A., Di Francesco F.

Affiliazione autori: Univ Pisa, Dept Chem & Ind Chem, Via Moruzzi 13, I-56124 Pisa, Italy;‎ CNR, Inst Clin Physiol, Via Moruzzi 1, I-56124 Pisa, Italy; CNR, Natl Inst Opt, Via Moruzzi 1, I-56124 Pisa, Italy; Univ Pisa, Res Ctr E Piaggio, Largo Lazzarino 1, I-56122 Pisa, Italy

Abstract: Objectives. -We developed a wearable sensor for the real time measurement of sweat rate in localized areas of the human body. This sensor represents the first step in the development of a wearable sensor network capable to estimate the global sweat rate via an ad hoc algorithm. Such device would be used to monitor athletes’ hydration status during training and improve their performances.
Equipment and methods. – For this study, we tested our sensor on thirteen football players during a cycling test on a cycle ergometer. The sweat rate sensor was compared to a medical device that, although measuring a different physiological process, provides discrete data based on the same working principle, i.e. the diffusion of the water vapour emitted from the skin.
Results. -Our sensor has a working range up to 400 g/m(2) .h. The statistical analysis and the Bland-Altman plot proved that our sensor is comparable to the medical device used as gold standard. At low sweat rate, the bias is 3.4 g/m(2) .h with a standard deviation of 7.6 g/m(2) .h. At maximum sweat rates, the bias is 2.3 g/m(2) .11 with a standard deviation 6.9 g/m(2) .11. The Pvalues for the Bland-Altman plots at low and maximum sweat rate (0.1331 and 0.2477 obtained by Kolmogorov-Smimov test, respectively) allow the hypothesis that there is a significant difference between our sweat rate sensor and the medical device to be rejected.
Conclusion. – We presented a prototype of a wearable sweat rate sensor for localized measurements. The trials on thirteen athletes proved that the performance of our sensor is comparable to that of a commercial medical device. This sweat rate sensor can provide valuable information on athletes’ hydration status. (C) 2017 Elsevier Masson SAS. All rights reserved.

Giornale/Rivista: SCIENCE & SPORTS

Volume: 33 (2)      Da Pagina: E51  A: E58

Maggiori informazioni: This work was partially supported by the EU-funded FP7 ICT-317894 SWAN-iCare project.
Parole chiavi: Sweat rate; Sports performance; Personalized training; Wearable sensor; Hydration